Evaluation of a Rebreathing System for Use with Portable Mechanical Ventilators

$43.00
Qty:

Blakeman T, Smith M, Branson R 99(5). 0 (Journal Article)

Introduction: Maximizing the capabilities of available lowflow oxygen is key to providing adequate oxygen to prevent/treat hypoxemia and conserve oxygen. We designed a closed-circuit system that allows rebreathing of gases while scrubbing carbon dioxide (CO2) in conjunction with portable mechanical ventilators in a bench model. Methods: We evaluated the system using two portable mechanical ventilators currently deployed by the Department of Defense-Zoll 731 and AutoMedx SAVe II-over a range of ventilator settings and lung models, using 1 and 3L/min low-flow oxygen into a reservoir bag. We measured peak inspired oxygen concentration (FiO2), CO2-absorbent life, gas temperature and humidity, and the effect of airway suctioning and ventilator disconnection on FiO2 on ground and at altitude. Results: FiO2 was ≥0.9 across all ventilator settings and altitudes using both oxygen flows. CO2-absorbent life was >7 hours. Airway humidity range was 87%-97%. Mean airway temperature was 25.4°C (SD 0.5°C). Ten-second suctioning reduced FiO2 22%-48%. Thirtysecond ventilator disconnect reduced FiO2 29%-63% depending on oxygen flow used. Conclusion: Use of a rebreathing system with mechanical ventilation has the potential for oxygen conservation but requires diligent monitoring of inspired FiO2 and CO2 to avoid negative consequences.